SELFCHECK Pregnancy Blood Test - Data Sheet for HCPs

PRODUCT: SELFCHECK PREGNANCY BLOOD TEST (hCG)

PRODUCT CODE: SCPBT1

REVISION DATE: 16th July 2025

Test Principle and Specifications

Intended use

The SELFCHECK Pregnancy Blood Test is a manual, qualitative, solid-phase immunochromatographic (lateral flow) immunoassay for the detection of human chorionic gonadotrophin (hCG) in a finger-prick whole blood sample.

Clinical Context and Limitations

This early pregnancy test can be used at any time of the day as it is largely unaffected by state of hydration compared to urine based pregnancy tests. The test uses a pair of antibodies specific for hCG. A positive result is indicated by two pink/red lines on the test cassette, and a negative result by one line.

The test may show a false positive result if the patient is taking drugs containing hCG, or drugs used in combination with hCG, for example; Pregnyl, Porfasi or Pergonal. A false positive result may occur if the patient has had a miscarriage, abortion or has given birth within the past 8 weeks, due to residual, but detectable hCG in the blood.

All medical tests, whether used in a hospital laboratory or at home, have limitations and carry a possibility of false positive or false negative results due to individual physiological variations and interpretation of the instructions for use (IFU). This test should be used as one part of a broader approach to health management, including healthcare professionals.

The test is not suitable for patients with haemchromatosis.

Performance Characteristics

Biomarker

hCG

Methodology

Immunochromatographic (lateral flow) assay

Sample type

Finger-prick whole blood sample

International standard

Calibrated to WHO 99/688

Cut-off level

25 mIU/mL

Performance vs. laboratory reference method (VIDAS)

Sensitivity: 100%

Specificity: 100%

Overall agreement: 100%

Regulatory and Quality Assurance

Device Classification

In Vitro Diagnostic (IVD) medical device for self-testing, and compliant with the UK Medical Devices Regulations 2002 (as amended).

Conformity Mark

CE marked for self-test use according to the EU In Vitro Diagnostic Directive 98/79 EC (IVDD), including the four-digit number (0483) of the responsible Notified Body. This number is a critical indicator that the device has undergone third-party review specifically for self-test use, distinguishing it from professional-use-only devices used in hospital laboratories and health care professionals. Notified and Approved Bodies are independent, third-party organisations designated by the Competent Authority (MHRA in the UK) to assess the conformity of IVDs to relevant regulations and standards including the IVDD and the In Vitro Diagnostic Regulations.

Regulatory Authority

Registered with the UK Medicines and Healthcare products Regulatory Agency (MHRA).

Quality Management System

Components are assembled in the United Kingdom by J R Biomedical Ltd., at an ISO 13485:2016 certified medical device manufacturing facility, independently audited by UKAS accredited auditors, for the design, development, and manufacture of in-vitro diagnostics. J R Biomedical Ltd (est. 2005), is a developer and supplier of diagnostics to the NHS and medical research laboratories.

Legal Manufacturer

VEDALAB, Rue de l'Expansion, Alençon, France.

UK Distributor

1st Health Products Ltd.

Further information

J R Biomedical Ltd. OpTIC Technology Centre, Wrexham University, St Asaph Business Park, St Asaph, LL17 0JD. United Kingdom.

Telephone 01745 535 169

https://jrbiomed.com